CRCST Chapter 6 Quiz (9th Edition): 20 Free Questions
Published 2026-11-09 · Facts and text checked 2026-11-02 · Questions from our CRCST study guide · Reviewed by Hana N., Quality Reviewer
This free quiz has 20 questions on the topics of chapter 6, Regulations and Standards, of the HSPA Sterile Processing Technical Manual, 9th edition: the regulations and standards that govern sterile processing. They are our own questions, from our CRCST study guide, not taken from the manual. Click an answer to read the explanation.
Our own questions on the topics of chapter 6, Regulations and Standards, of the HSPA Sterile Processing Technical Manual, 9th edition. On the exam this is part of Departmental Considerations. Open each answer to read the explanation.
Question 1Medium
A technician is cleaning a surgical instrument with visible blood and tissue. According to AAMI ST79, what is the recommended first step in the cleaning process?
Show answer and explanation
Correct answer: A. Rinse the instrument with cold water to remove gross debris.
AAMI ST79 recommends that cleaning begin with removal of gross soil, typically by rinsing with cold water to prevent protein fixation. Hot water can coagulate proteins, making cleaning harder. Disinfectant and enzymatic foam are used after initial cleaning, and ultrasonic cleaning is a later step.
From practice test 5 of our CRCST study guide (question 77). Original question, not an official exam item.
Question 2Easy
After a needlestick injury in the decontamination area, what must the employer provide under OSHA's Bloodborne Pathogens Standard?
Show answer and explanation
Correct answer: D. Immediate medical evaluation and post-exposure prophylaxis (PEP) at no cost.
OSHA 1910.1030 requires employers to offer a confidential medical evaluation and follow-up, including post-exposure prophylaxis, immediately after a needlestick. The other options are insufficient; a tetanus shot may be indicated but is not the primary OSHA requirement, and paid leave is not mandated.
From practice test 5 of our CRCST study guide (question 19). Original question, not an official exam item.
Question 3Easy
Which labeling standard is legally required by OSHA for chemical containers in the central service department, and may coexist with other systems such as NFPA 704 or HMIS?
Show answer and explanation
Correct answer: D. GHS labels are the legal standard; NFPA/HMIS may be used in addition.
OSHA's Hazard Communication Standard mandates GHS labels on all chemical containers. NFPA 704 and HMIS are complementary systems that may appear on containers but are not substitutes for the required GHS label.
From practice test 2 of our CRCST study guide (question 52). Original question, not an official exam item.
Question 4Easy
Under the OSHA Bloodborne Pathogens Standard, an employer must offer the hepatitis B vaccination to an at-risk central service technician:
Show answer and explanation
Correct answer: B. Within 10 days of initial assignment to tasks with occupational exposure
The standard mandates that the vaccination be offered within 10 days of assignment to tasks with exposure risk. It is not conditional on injury, not annual, and must be at no cost to the employee.
From practice test 2 of our CRCST study guide (question 127). Original question, not an official exam item.
Question 5Medium
Under OSHA's Bloodborne Pathogens Standard, which of the following is an employer requirement for central service employees who may have occupational exposure?
Show answer and explanation
Correct answer: C. Provide personal protective equipment at no cost to the employee
OSHA's Bloodborne Pathogens Standard requires employers to provide PPE at no cost. Air quality monitoring is not a standard requirement; sterilization logs are under Joint Commission/OSHA record retention but not this standard; hepatitis B vaccination must be offered within 10 days of assignment, not 30 days after exposure.
From practice test 2 of our CRCST study guide (question 115). Original question, not an official exam item.
Question 6Medium
During a Joint Commission survey, the surveyor asks why the sterilization load record includes both 'Time In' and 'Time Out' fields. Which response best explains the rationale?
Show answer and explanation
Correct answer: B. To provide a complete chronological record of the sterilization process
The time in and time out fields document the start and end of the sterilization cycle, providing a chronological record. While exposure time is important, it is derived from the difference between these times, but the primary purpose is recording the entire cycle timeline.
From practice test 7 of our CRCST study guide (question 135). Original question, not an official exam item.
Question 7Medium
A surveyor asks a central service manager for evidence that the department follows nationally recognized guidelines for preventing infections related to reprocessed medical devices. Which Joint Commission standard is the surveyor most likely referencing?
Show answer and explanation
Correct answer: D. IC.02.02.01
Joint Commission standard IC.02.02.01 specifically addresses preventing infections related to medical devices and expects facilities to follow nationally recognized guidelines, such as AAMI standards. The other options cover leadership, environment of care, and human resources, respectively.
From practice test 2 of our CRCST study guide (question 24). Original question, not an official exam item.
Question 8Easy
During a Joint Commission survey, a hospital is cited for incomplete sterilization records. What is the term for this citation?
Show answer and explanation
Correct answer: B. A Requirement for Improvement (RFI).
The Joint Commission issues an RFI for noncompliance that requires a written plan of correction. It is not a recommendation or an immediate suspension; it triggers a follow-up survey to verify corrective actions.
From practice test 5 of our CRCST study guide (question 58). Original question, not an official exam item.
Question 9Hard
A new cleaning chemical is introduced in central service. Under the OSHA Hazard Communication Standard, which of the following actions must the employer take?
Show answer and explanation
Correct answer: A. Ensure the chemical’s safety data sheet is available within seconds in the department
The Hazard Communication Standard requires that safety data sheets be readily accessible to employees at all times. Warning signs are not the primary requirement; respirator fit tests are only needed if respirators are required; submission to fire department is not part of this standard.
From practice test 2 of our CRCST study guide (question 143). Original question, not an official exam item.
Question 10Medium
A central service technician notices a new disinfectant in the work area with a GHS label but no training has been provided about the new label elements. Which OSHA standard is being violated?
Show answer and explanation
Correct answer: C. The failure to train employees on the new hazard labels before first use
OSHA's Hazard Communication Standard requires employers to train employees on the new GHS label elements before they are first exposed to the chemical. Even if the label itself is correct, without training, the employer is noncompliant.
From practice test 2 of our CRCST study guide (question 22). Original question, not an official exam item.
Question 11Easy
According to OSHA regulations, how long must records of employee exposure to hazardous materials be retained?
Show answer and explanation
Correct answer: D. 30 years
OSHA's Hazard Communication Standard (29 CFR 1910.1200) requires that exposure records be retained for at least 30 years. The seven-year option applies to sterilization records under Joint Commission standards, not OSHA.
From practice test 7 of our CRCST study guide (question 72). Original question, not an official exam item.
Question 12Medium
An OSHA inspector observes a central service decontamination area and notices that sharps disposal containers are not puncture-resistant. Which OSHA standard directly addresses the required construction of sharps containers as an engineering control?
Show answer and explanation
Correct answer: A. Bloodborne Pathogens Standard (29 CFR 1910.1030)
The Bloodborne Pathogens Standard mandates that sharps containers be puncture-resistant, leakproof, and closable as engineering controls. The other standards address chemical labels, PPE selection, and injury recording, respectively.
From practice test 2 of our CRCST study guide (question 110). Original question, not an official exam item.
Question 13Medium
After receiving a Requirement for Improvement (RFI) for documentation gaps, what is the Joint Commission's typical next step?
Show answer and explanation
Correct answer: D. Conducting a follow-up survey to verify corrective actions.
The Joint Commission requires a written plan of correction and then conducts a follow-up survey to ensure the deficiencies have been resolved. Fines and immediate revocation are not standard practices for a single RFI.
From practice test 5 of our CRCST study guide (question 23). Original question, not an official exam item.
Question 14Medium
A manufacturer wants to market a new surgical instrument that is substantially equivalent to a device already on the market. Which FDA process is most likely required before marketing?
Show answer and explanation
Correct answer: D. 510(k) premarket notification clearance.
Devices found to be substantially equivalent to a predicate device can be cleared via the 510(k) route. PMA is for novel, high-risk devices; IDE is for investigational devices; QSR is a quality management requirement but does not replace market clearance.
From practice test 3 of our CRCST study guide (question 81). Original question, not an official exam item.
Question 15Hard
A central service storage area contains several gallon containers of alcohol-based disinfectant and an ethylene oxide (EtO) sterilizer canister. During a Joint Commission environment of care survey, which requirement must the facility demonstrate compliance with?
Show answer and explanation
Correct answer: B. The area must have a sprinkler system and ventilation meeting NFPA codes
The Joint Commission enforces NFPA codes for hazardous material storage, including sprinklers and ventilation for alcohol and EtO. Locked cabinets are not specified; open shelving is not recommended; the distance from heat is an OSHA detail, not the primary environmental standard.
From practice test 2 of our CRCST study guide (question 149). Original question, not an official exam item.
Question 16Medium
A biological indicator from a steam sterilization load shows growth after incubation. According to AAMI standards, which of the following is the correct immediate action?
Show answer and explanation
Correct answer: B. Quarantine the load, recall any distributed items, and remove the sterilizer from service for investigation.
AAMI ST79 specifies that a positive biological indicator requires immediate quarantine of the load, recall of items, and removal of the sterilizer from service for cause analysis. Running a second BI on the same load (option A) delays action and is not recommended; the load is already suspect.
From practice test 3 of our CRCST study guide (question 6). Original question, not an official exam item.
Question 17Easy
A gravity displacement steam sterilizer is set to run a wrapped instrument cycle at 250°F (121°C). What are the standard cycle parameters for exposure time and chamber pressure?
Show answer and explanation
Correct answer: A. 30 minutes exposure, 15-17 psi
For a gravity displacement cycle at 250°F, AAMI ST79 specifies a minimum exposure time of 30 minutes and chamber pressure of 15-17 psi.
From practice test 2 of our CRCST study guide (question 31). Original question, not an official exam item.
Question 18Easy
According to AAMI standards, what is the requirement for biological indicators (BIs) in sterilization loads containing implantable devices?
Show answer and explanation
Correct answer: D. A BI must be placed in every implant load and the load cannot be released until the BI result is known.
AAMI ST79 requires that every load containing an implantable device must be monitored with a biological indicator, and the implant cannot be released until the BI result is confirmed negative. The other options are incorrect because BI monitoring is required for all implant loads regardless of cycle type or other indicator results.
From practice test 3 of our CRCST study guide (question 143). Original question, not an official exam item.
Question 19Medium
According to AAMI ST79, which of the following is a correct guideline for load configuration in a steam sterilizer?
Show answer and explanation
Correct answer: A. Stack peel pouches on their edges to prevent condensation
AAMI ST79 recommends placing peel pouches on their edges to allow air removal and steam contact, preventing moisture accumulation. Trays should be on edge, not flat; container lids should be closed with valves open per IFU; textured surfaces do not affect sterilization.
From practice test 4 of our CRCST study guide (question 25). Original question, not an official exam item.
Question 20Medium
A manufacturer has developed a novel implantable orthopedic device. Under FDA classification, this device would most likely be:
Show answer and explanation
Correct answer: B. Class III, because it supports life or presents unreasonable risk.
Implantable devices that support life or present unreasonable risk are typically Class III, requiring premarket approval. Class I and II are lower risk categories.
From practice test 6 of our CRCST study guide (question 72). Original question, not an official exam item.
0 of 20 answered. Your result shows each area against our study target.
Want more mixed questions? Go back to the CRCST practice test hub.
Want the full CRCST preparation?
These questions come from our CRCST study guide: full-length practice tests with every answer explained, flashcards, an AI study tutor and the audiobook.