CRCST Chapter 15 Quiz (9th Edition): 20 Free Questions

Published 2026-11-15 · Facts and text checked 2026-11-02 · Questions from our CRCST study guide · Reviewed by Hana N., Quality Reviewer

This free quiz has 20 questions on the topics of chapter 15, Low-Temperature Sterilization, of the HSPA Sterile Processing Technical Manual, 9th edition: ethylene oxide, hydrogen peroxide and other low-temperature methods. They are our own questions, from our CRCST study guide, not taken from the manual. Click an answer to read the explanation.

Our own questions on the topics of chapter 15, Low-Temperature Sterilization, of the HSPA Sterile Processing Technical Manual, 9th edition. On the exam this is part of Sterilization Process. Open each answer to read the explanation.

Question 1Easy

Which microorganism is used as the biological indicator for ethylene oxide sterilization?

Show answer and explanation

Correct answer: D. Bacillus atrophaeus

Bacillus atrophaeus spores are the standard biological indicator for EtO sterilization because of their resistance to the gas. Geobacillus stearothermophilus is used for steam sterilization.

From practice test 5 of our CRCST study guide (question 14). Original question, not an official exam item.

Question 2Easy

Which of the following correctly lists the three phases of an ethylene oxide sterilization cycle in order?

Show answer and explanation

Correct answer: C. Preconditioning, sterilization exposure, aeration

The cycle begins with preconditioning (temperature and humidity stabilization), then sterilization exposure (gas contact), and finally aeration (gas removal).

From practice test 2 of our CRCST study guide (question 64). Original question, not an official exam item.

Question 3Easy

What is the minimum personal protective equipment required when removing sterilized items from an EtO sterilizer after aeration is complete?

Show answer and explanation

Correct answer: B. Chemical-resistant gloves and eye protection

After aeration, residual EtO is low but chemical-resistant gloves and eye protection are recommended to prevent dermal or ocular exposure.

From practice test 3 of our CRCST study guide (question 116). Original question, not an official exam item.

Question 4Medium

Which action is an example of a work practice control for EtO sterilization safety?

Show answer and explanation

Correct answer: B. Verifying the door seal integrity before each cycle.

Work practice controls are procedures performed by staff, such as door seal verification. Option A is an engineering control, C is PPE, D is not specific.

From practice test 3 of our CRCST study guide (question 103). Original question, not an official exam item.

Question 5Hard

A central service department processes implantable devices using ethylene oxide. After aeration, which action is the most appropriate to confirm that residual EtO levels are safe for patient use?

Show answer and explanation

Correct answer: A. Use a gas chromatograph to test a representative sample from the load.

Gas chromatography directly measures residual EtO levels, providing quantitative data for implantable devices where safety margins are critical. Biological indicators test sterilization efficacy, not residual gas.

From practice test 3 of our CRCST study guide (question 137). Original question, not an official exam item.

Question 6Easy

According to AAMI standards, which of the following must be maintained for each type of load processed in an ethylene oxide sterilizer?

Show answer and explanation

Correct answer: B. Written instructions from the manufacturer specifying the sterilization parameters

AAMI standards require that sterile processing facilities maintain written instructions for each load type, including manufacturer's parameters. Biological indicator logs are important but not the requirement per load type; Bowie-Dick tests are for prevacuum steam sterilizers, not EtO.

From practice test 4 of our CRCST study guide (question 58). Original question, not an official exam item.

Question 7Hard

A technician is setting up an EtO cycle for a load of heat-sensitive catheters. The manufacturer's instructions for this load specify an exposure time of 3 hours at a gas concentration of 600 mg/L. If the sterilizer can only deliver 450 mg/L, how should the technician adjust the cycle to maintain sterility assurance?

Show answer and explanation

Correct answer: D. Increase the exposure time beyond 3 hours, as lower concentration requires longer exposure.

Standard EtO concentration range is 450-1200 mg/L. Lower concentration within that range requires longer exposure to achieve the same sterility assurance level. The manufacturer's instructions must be followed, but if concentration is reduced, time must be increased proportionally.

From practice test 4 of our CRCST study guide (question 41). Original question, not an official exam item.

Question 8Easy

In ethylene oxide sterilization, the concentration of the gas is typically expressed in which unit?

Show answer and explanation

Correct answer: A. Milligrams per liter.

Standard EtO concentration is measured in milligrams per liter (mg/L). Parts per million may be used for safety monitoring, but mg/L is the unit for sterilization parameters.

From practice test 7 of our CRCST study guide (question 85). Original question, not an official exam item.

Question 9Medium

An EtO cycle completes and the sterilizer display indicates 'cycle finished.' Why must the door interlock remain engaged?

Show answer and explanation

Correct answer: A. To prevent release of residual EtO gas during the aeration phase

The door interlock stays locked until aeration is complete to contain residual EtO, which is hazardous. Opening the door early would expose staff to harmful gas and compromise safety.

From practice test 2 of our CRCST study guide (question 80). Original question, not an official exam item.

Question 10Hard

A maintenance technician needs to replace the activated carbon exhaust filter on an EtO sterilizer. According to safety protocols, which level of PPE is required?

Show answer and explanation

Correct answer: A. Full PPE including a respirator

Filter changes involve potential exposure to high concentrations of trapped EtO, so full PPE with a respirator is mandatory. Lesser PPE is insufficient.

From practice test 3 of our CRCST study guide (question 135). Original question, not an official exam item.

Question 11Medium

A busy central service department runs multiple EtO cycles per day. To maximize sterilizer utilization, they should:

Show answer and explanation

Correct answer: D. Transfer the load to a separate mechanical aeration chamber after exposure

Separate aeration chambers allow the sterilizer to begin the next cycle immediately, increasing throughput. Combined chambers tie up the sterilizer during aeration; reducing aeration time is unsafe; extending exposure does not affect aeration.

From practice test 6 of our CRCST study guide (question 33). Original question, not an official exam item.

Question 12Medium

A new technician asks why the door of an EtO sterilizer should never be opened during the cycle, even briefly. Which is the primary reason?

Show answer and explanation

Correct answer: A. The gas is toxic and would be released into the room.

The primary safety concern is the release of ethylene oxide, a known carcinogen and toxic gas, into the work environment. While flammability is also a hazard, toxicity is the immediate and continuous risk during the cycle.

From practice test 5 of our CRCST study guide (question 41). Original question, not an official exam item.

Question 13Hard

A central service department receives a new heat-sensitive surgical device that requires EtO sterilization. The device manufacturer did not provide written sterilization instructions. What is the appropriate action according to AAMI standards?

Show answer and explanation

Correct answer: D. Contact the device manufacturer to obtain validated written instructions before processing.

AAMI standards require that written manufacturer instructions be followed exactly. Without them, the facility cannot verify that the cycle parameters are adequate for that specific device. Processing without validated instructions risks sterilization failure.

From practice test 4 of our CRCST study guide (question 22). Original question, not an official exam item.

Question 14Hard

A facility processes heat-sensitive items using EtO at 55°C. Which exposure time range is most appropriate for this cycle?

Show answer and explanation

Correct answer: B. 1-2 hours

At 55°C, standard EtO exposure is 1 to 2 hours because higher temperature accelerates microbial kill. The other options are either too short for sterility assurance or correspond to lower temperature cycles.

From practice test 5 of our CRCST study guide (question 70). Original question, not an official exam item.

Question 15Hard

A new technician suggests using a multi-purpose room for the EtO sterilizer to save space. What is the primary safety concern with this proposal?

Show answer and explanation

Correct answer: B. The room lacks continuous gas monitoring and dedicated ventilation

EtO sterilizers must be in a dedicated room with continuous gas monitoring and proper ventilation to protect staff from exposure. Using a multi-purpose room without these controls creates an unacceptable safety hazard.

From practice test 2 of our CRCST study guide (question 41). Original question, not an official exam item.

Question 16Hard

After an ethylene oxide sterilization cycle, staff question why aeration is necessary before the load can be used. Which response best explains the reason?

Show answer and explanation

Correct answer: B. Aeration allows residual EtO absorbed into plastics and rubber to desorb.

EtO absorbs into materials like plastics and rubber; aeration desorbs the gas to safe levels. While aeration may also remove moisture, the primary reason is to eliminate toxic residuals. Cooling is secondary, and indicators are unrelated.

From practice test 7 of our CRCST study guide (question 61). Original question, not an official exam item.

Question 17Easy

During the preconditioning phase of an ethylene oxide sterilization cycle, which of the following must be verified before gas injection begins?

Show answer and explanation

Correct answer: D. Temperature and humidity are stable within acceptable ranges.

Preconditioning requires stable temperature and humidity to ensure effective microbial kill; gas injection should only commence after these parameters are verified. The other options are unrelated to preconditioning: pressure is a separate parameter, vacuum leak testing is for prevacuum cycles, and aeration occurs after sterilization.

From practice test 7 of our CRCST study guide (question 89). Original question, not an official exam item.

Question 18Medium

A central service department uses a separate aeration chamber for EtO-sterilized items. What is the primary advantage of this arrangement?

Show answer and explanation

Correct answer: B. It allows the sterilizer to process another load while aeration occurs

A separate aeration chamber frees the sterilizer for the next cycle, increasing throughput. The other options do not describe the benefit of separate chambers.

From practice test 2 of our CRCST study guide (question 18). Original question, not an official exam item.

Question 19Medium

After loading an EtO sterilizer and prior to starting the cycle, a technician notices a small gap in the door gasket. According to standard work practices, what should the technician do?

Show answer and explanation

Correct answer: D. Contact maintenance to replace the gasket before proceeding.

Verifying door seal integrity is a required work practice before each cycle; a damaged gasket must be replaced to ensure containment of toxic gas. Relying on vacuum or using temporary fixes compromises safety and is not permitted.

From practice test 5 of our CRCST study guide (question 87). Original question, not an official exam item.

Question 20Easy

If the mechanical aeration system for an EtO sterilizer fails and items must be aerated at room temperature (approximately 22°C), what is the standard aeration time range?

Show answer and explanation

Correct answer: D. 7-14 days

Room temperature aeration for EtO takes 7-14 days because desorption of the gas from materials is much slower at lower temperatures.

From practice test 4 of our CRCST study guide (question 96). Original question, not an official exam item.