CBSPD Roles and Responsibilities Quiz: 20 Free Questions
Published 2026-11-11 · Facts and text checked 2026-11-02 · Questions from our CBSPD study guide · Reviewed by Hana N., Quality Reviewer
This free CBSPD Roles and Responsibilities Quiz has 20 questions from the practice tests in our CBSPD study guide, on standards, quality and the role of the technician. Click an answer to see if it is right and read the explanation, then check your result against our study target.
Standards, quality and the role of the technician. On the exam this is part of Roles and Responsibilities (13%). Open each answer to read the explanation.
Question 1Medium
A technician needs to run the Bowie-Dick test for a pre-vacuum steam sterilizer. When is the test performed and what does it confirm?
Show answer and explanation
Correct answer: A. Daily, in an empty chamber, before the first cycle; confirms complete air removal
AAMI ST79 requires the Bowie-Dick test to be run daily in an empty pre-vacuum sterilizer before the first processing cycle. It verifies complete air removal, not sterilant penetration or microbial kill.
From practice test 8 of our CBSPD study guide (question 47). Original question, not an official exam item.
Question 2Hard
A surgical team requests immediate-use steam sterilization for a set of instruments needed for a scheduled total joint replacement, because the sterile set was accidentally dropped and cannot be used. Which response is most consistent with current guidance on immediate-use steam sterilization?
Show answer and explanation
Correct answer: B. Decline the request for this scheduled case, since immediate-use steam sterilization is reserved for urgent clinical situations and is not a routine substitute for standard reprocessing.
Immediate-use steam sterilization is limited to occasions when no alternative exists and the clinical need is urgent; it is not intended for scheduled cases or for anything that can be conventionally reprocessed. The dropped set should be reprocessed through the routine wrapping and sterilization process.
From practice test 2 of our CBSPD study guide (question 118). Original question, not an official exam item.
Question 3Medium
A rigid laparoscopic cannula with a narrow lumen is being cleaned before sterilization. Which cleaning action is correct per AAMI ST90?
Show answer and explanation
Correct answer: B. Flush the lumen with detergent solution and pass a brush sized to the lumen diameter through both openings.
Lumens require active cleaning: the brush must match the lumen diameter and be passed through both ends, and the lumen must be flushed with detergent solution. A standing ultrasonic soak or saline rinse does not mechanically remove interior soil.
From practice test 6 of our CBSPD study guide (question 18). Original question, not an official exam item.
Question 4Medium
A rigid container system is being prepared to hold a set of instruments for sterilization. Which condition makes the container an acceptable packaging system?
Show answer and explanation
Correct answer: C. The rigid container must be compatible with the sterilization method, with its filters and valves inspected according to the manufacturer's instructions.
AAMI ST90 and the associated container standards require that the container be compatible with the sterilization method and that filters, valves, and gaskets be inspected per the manufacturer's instructions. A missing filter or loosely secured lid would compromise the sterile barrier.
From practice test 6 of our CBSPD study guide (question 38). Original question, not an official exam item.
Question 5Hard
A heat-sensitive device cannot tolerate steam, and its manufacturer's IFU specifies a low-temperature sterilization method. Which action is correct?
Show answer and explanation
Correct answer: C. Process the device using the low-temperature sterilization method specified in its IFU, with the full validated parameters.
When a device cannot tolerate steam, the manufacturer's validated low-temperature sterilization method, including exposure time and temperature, must be followed. Shortening a steam cycle or using high-level disinfection instead of sterilization does not meet the validated process.
From practice test 6 of our CBSPD study guide (question 47). Original question, not an official exam item.
Question 6Easy
A sterile processing technician is preparing a case cart for a scheduled surgical procedure. Which step follows the AORN guideline on case cart preparation?
Show answer and explanation
Correct answer: B. Use the procedure-specific case cart list to arrange and cover the cart before transporting it to the OR.
AORN guidelines direct that case carts be assembled according to a procedure-specific list and be covered or closed during transport to protect the contents. An uncovered cart exposes sterile supplies to contamination.
From practice test 6 of our CBSPD study guide (question 34). Original question, not an official exam item.
Question 7Medium
A biological indicator from a steam sterilization load returns positive. What should be done with all items processed in that load?
Show answer and explanation
Correct answer: D. Recall the load and quarantine all items
A positive biological indicator is a potential patient safety event, so the load must be recalled and all items contained. Releasing, rerunning, or returning items ignores the need to explain the failure.
From practice test 3 of our CBSPD study guide (question 69). Original question, not an official exam item.
Question 8Medium
During a process audit, a technician is observed skipping the verification of one instrument. What is the appropriate immediate use of this observation?
Show answer and explanation
Correct answer: C. Review the finding with the technician to identify where the process breaks, without making it a disciplinary record.
Audit observations are intended to locate process failures, not to punish individuals. Reviewing the finding with the technician reveals whether the skip was caused by a confusing instruction or a missing checkpoint.
From practice test 4 of our CBSPD study guide (question 41). Original question, not an official exam item.
Question 9Hard
A positive biological indicator is reported from a steam sterilizer, and the sterilizer is taken out of service. Which condition must be met before the sterilizer can be returned to service?
Show answer and explanation
Correct answer: A. A root cause analysis must explain the failure
The sterilizer remains out of service until a root cause analysis explains the failure; running test cycles or cleaning the storage room does not identify why the biological indicator was positive.
From practice test 3 of our CBSPD study guide (question 135). Original question, not an official exam item.
Question 10Medium
A technician notices that the department's written policy for steam sterilizer loading contradicts the current ANSI/AAMI ST79 standard that the policy was supposed to implement. Which action is correct?
Show answer and explanation
Correct answer: D. Follow the AAMI standard until the policy is corrected.
The AAMI standard is the authoritative reference for the reprocessing process, so a conflicting departmental policy must be corrected and the standard followed in the interim. The same principle applies when a manufacturer's IFU conflicts with a general policy.
From practice test 4 of our CBSPD study guide (question 33). Original question, not an official exam item.
Question 11Medium
A scrub person calls sterile processing to say a contaminated grasper from the operating room was set aside on a back table and is now dry. Which practice should have been followed to make transport safe and cleaning effective?
Show answer and explanation
Correct answer: B. Keep the instrument moist, remove gross soil, and place it in a closed container for transport to decontamination.
AORN requires that contaminated items be transported in a closed, covered container and kept moist when decontamination is delayed. An open basin or dry transport allows soil to dry and creates a contamination risk.
From practice test 9 of our CBSPD study guide (question 95). Original question, not an official exam item.
Question 12Medium
A flexible endoscope has completed manual cleaning. Per the department's cleaning verification policy, what must happen before the scope is placed into the high-level disinfectant?
Show answer and explanation
Correct answer: A. The scope must pass a protein residue test.
After manual cleaning, a flexible endoscope must pass a protein residue test before high-level disinfection. This verifies that cleaning achieved the required level.
From practice test 9 of our CBSPD study guide (question 85). Original question, not an official exam item.
Question 13Easy
A quality assurance schedule for daily checks in a sterile processing department should include which of the following elements for each check?
Show answer and explanation
Correct answer: C. A sign-off line with a date and an initials box
Each scheduled check must have a sign-off line with a date and initials box so that verification is documented. The other choices describe documentation for other purposes, not the required sign-off element.
From practice test 5 of our CBSPD study guide (question 107). Original question, not an official exam item.
Question 14Hard
A routine biological indicator placed in a steam sterilizer produces a positive result after incubation. What is the most appropriate response?
Show answer and explanation
Correct answer: C. Treat the load as nonsterile, remove the sterilizer from service, recall items processed since the last negative biological indicator, and investigate the cause before resuming use.
A positive biological indicator means the sterilization cycle did not kill the test spores, so the load cannot be considered sterile. The proper response is to remove the sterilizer from service, recall all loads since the last acceptable biological indicator, and identify the cause before returning it to use.
From practice test 2 of our CBSPD study guide (question 6). Original question, not an official exam item.
Question 15Hard
After a pre-vacuum cycle, the chemical indicator on a wrapped tray has turned to its endpoint color, but the biological indicator test result is positive. What should the technician conclude?
Show answer and explanation
Correct answer: C. The tray cannot be released; a positive biological indicator means the items are not considered sterile
A biological indicator is the only monitoring device that directly verifies microbial kill, and a positive result means the load cannot be considered sterile. A chemical indicator only demonstrates that the item was exposed to the sterilant, so its endpoint color does not override a positive biological indicator.
From practice test 8 of our CBSPD study guide (question 21). Original question, not an official exam item.
Question 16Medium
A supervisor suspects that the written count-sheet procedure itself is outdated for a newly introduced instrument set. Which type of audit is most appropriate to determine whether the procedure, rather than staff performance, is the problem?
Show answer and explanation
Correct answer: B. Product audit, because it compares the set contents against the manufacturer's instructions for use.
A product audit compares the assembled instrument set against the manufacturer's instructions for use. If the set does not match the IFU even when staff follow the count sheet exactly, the count-sheet procedure itself is outdated; a process audit would only confirm that staff followed the written steps.
From practice test 4 of our CBSPD study guide (question 85). Original question, not an official exam item.
Question 17Easy
A sterile processing manager must validate a new ethylene oxide sterilization cycle and needs a standard devoted entirely to this sterilization method. Which standard should be used?
Show answer and explanation
Correct answer: A. ISO 11135
ISO 11135 is the standard dedicated to development, validation, and routine control of ethylene oxide sterilization of medical devices. ST79 addresses steam, while ISO 13485 addresses quality management systems.
From practice test 9 of our CBSPD study guide (question 42). Original question, not an official exam item.
Question 18Medium
A biological indicator (BI) from an implant load tests positive. Which action must be taken first?
Show answer and explanation
Correct answer: A. Take the sterilizer out of service and recall or quarantine the affected load
A positive biological indicator is a potential patient safety event, so the sterilizer must be taken out of service and all potentially affected items must be recalled or quarantined. Rerunning the cycle or releasing the load does not address the failure.
From practice test 2 of our CBSPD study guide (question 9). Original question, not an official exam item.
Question 19Easy
According to AAMI ST79, which temperature range is recommended for a sterile storage room?
Show answer and explanation
Correct answer: A. 18°C to 24°C (65°F to 75°F)
AAMI ST79 specifies 18°C to 24°C (65°F to 75°F) as the recommended temperature range for sterile storage rooms. The other ranges fall outside the standard's requirements and could compromise package integrity or microbial control.
From practice test 4 of our CBSPD study guide (question 61). Original question, not an official exam item.
Question 20Easy
The upper acceptable temperature limit for a sterile storage room is:
Show answer and explanation
Correct answer: C. 24°C
The upper limit is 24°C; readings trending toward it indicate the cooling unit may be failing. Lower values are acceptable but are not the upper limit, and 26°C exceeds the standard.
From practice test 5 of our CBSPD study guide (question 128). Original question, not an official exam item.
0 of 20 answered. Your result shows each area against our study target.
Want more mixed questions? Go back to the CBSPD practice test hub.